Analysis of an Anti-Aliasing and Reconstruction System

Authors

  • Mashal Ahmad Electrical Engineering Department, University of South Australia, Adelaide, Australia Author
  • Zhong Wang Electrical Engineering Department, University of South Australia, Adelaide, Australia Author
  • Fakhar Sial Electrical Engineering Department, University of South Australia, Adelaide, Australia Author

DOI:

https://doi.org/10.57041/s4jfnh46

Keywords:

Electrocardiogram (ECG), low-pass filter, aliasing, noise

Abstract

This research develops a complete signal-processing chain for a digital electrocardiogram (ECG) monitor. The signal contains 20 Hz and 60 Hz cardiac components plus 150 Hz noise. A 6th-order Butterworth low‑pass filter with an 80 Hz cut-off removes the noise before sampling at 250 Hz, preventing aliasing. The system uses causal zero-order hold reconstruction and is proven stable, linear, and causal. MATLAB simulations validate accurate signal recovery with minimal distortion.