Analysis of an Anti-Aliasing and Reconstruction System
DOI:
https://doi.org/10.57041/s4jfnh46Keywords:
Electrocardiogram (ECG), low-pass filter, aliasing, noiseAbstract
This research develops a complete signal-processing chain for a digital electrocardiogram (ECG) monitor. The signal contains 20 Hz and 60 Hz cardiac components plus 150 Hz noise. A 6th-order Butterworth low‑pass filter with an 80 Hz cut-off removes the noise before sampling at 250 Hz, preventing aliasing. The system uses causal zero-order hold reconstruction and is proven stable, linear, and causal. MATLAB simulations validate accurate signal recovery with minimal distortion.Downloads
Published
2025-12-30
Issue
Section
Articles
License
Copyright (c) 2025 https://grsh.org/journal1/index.php/jaic/cr

This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.
How to Cite
Analysis of an Anti-Aliasing and Reconstruction System. (2025). Journal of Artificial Intelligence and Computing, 3(2), 7-10. https://doi.org/10.57041/s4jfnh46